AW00123409000 Physical InterfaceBasler ace USB 3.0 915.11 Configuring Input Lines and Signals5.11.1 Selecting an Input Line as the Source Signal for aCamera FunctionYou can select input line Line 1 and GPIO lines Line 3 and Line 4, if configured for input, to act asthe source signal for the following camera functions: Frame Burst Start trigger Frame Start trigger Counter 1 resetWhenever a proper electrical signal is applied to the selected line, the camera will recognize thesignal as signal for the selected camera function.For example, when Line 1 was selected to act as the source signal for the frame burst start trigger,camera will recognize an electrical signal applied to Line 1 as a frame burst start trigger.Note: When you apply an electrical signal to the input line the electrical signal must be appropriatelytimed for the function.For detailed information about selecting an input line to act as the source signal for the frame burst start trigger and for details about how the frame burst start trigger operates,see Section 6.3 on page 115. the frame start trigger and for details about how the frame start trigger operates, seeSection 6.4 on page 122. counter 1 reset and for details about how the counter value chunk feature operates, seeSection 7.23.3.6 on page 344.By default, input line Line 1 is selected as the source signal for the frame starttrigger.