CASMED 740 MONITOR SERVICE MANUAL21-02-0174 REV. 02 05/08103Installation1) Install the RS232/Nurse Call Interface Board into the Module case and secure it with thefour (4) screws previously removed.2) Reconnect the ribbon cable.3) Place the two case halves together being careful not to pinch the ribbon cable.4) Secure the two case halves together using the four (4) screws previously removed.5) From the Label Set, re-install the screw fastener covers and the module’s top label.REPLACING THE SURETEMP® PLUS TEMPERATURE BOARDRemoval1) Remove the one (1) screw that secures the Module to the monitor’s rear case.2) Remove the label from the top of the module.3) Remove the four (4) screw fastener covers.4) Remove the four (4) screws that secure the Module’s cases together.5) Separate the Module’s front and rear cases, being careful not to stress the internalribbon cable.6) Disconnect the ribbon cable from the SureTemp Plus Temperature Board.7) Remove the four (4) screws that secure the board to the case.8) Remove the board.Figure 24: Replacing the SureTemp Plus Temperature Board