CASMED 740 MONITOR SERVICE MANUAL21-02-0174 REV. 02 05/0812Monitor mode is normally used for longer term monitoring and when difficult situations preventaccurate temperature from being taken in the predictive mode. In monitoring mode, the probemust be in contact with tissue for at least three (3) minutes for accurate oral / rectal temperaturemeasurement and five (5) minutes for accurate axillary temperature measurement.The default setting used by the CASMED 740 Monitor for temperature determinations is thenormal (predictive) mode.NOTE:Axillary temperature readings may only be taken in the Neonate monitoring mode.PATIENT ENVIRONMENTThe CASMED 740 Monitor has been tested with specific parts of the “system” used within thePatient Environment. Figure 1, defines the Patient Environment.Figure 1: Patient EnvironmentThe parts of the CASMED 740 Monitor “system” that can be used in the Patient Environmentare defined as;The CASMED 740 MonitorAppropriate Accessories, listed in the ACCESSORIES section of the User’s ManualLine CordOptional RS232 / Nurse Call OptionCitizen CMP-10 Mobile PrinterRS232 Interconnect Cable (supplied with printer)AC Adapter / Charger, Model TRC-09-1100-M from Group West or equivalent (suppliedwith printer)Table 1: Parts of the System