CASMED 740 MONITOR SERVICE MANUAL21-02-0174 REV. 02 05/08559. ROUTINE MAINTENANCECLEANINGCLEANING OVERVIEWWARNING:Do not, under any circumstances, perform any testing or maintenance on the monitor while themonitor is being used to monitor a patient.CAUTION:Do not open the monitor to clean or repair it. Contact CAS Medical System for service needs.CAUTION:Unplug the monitor from the AC power source and remove all the accessories from the monitorbefore cleaning. The monitor must be turned off and not running on the internal battery. Neverclean the monitor when it is being operated.THE MONITOROn a daily basis, examine the monitor’s case for any damages and check the AC power cord forbent or broken prongs, cracks or fraying. Neither the monitor nor the power cord should beused if damaged. If any damage is noted, contact the appropriate service personnel.CAUTION:Do not spray any water or cleaning solution directly onto the monitor.As needed, clean the monitor using a soft cloth dampened with a mild dishwashing detergentsolution and gently rub the soiled area until clean. Use a clean soft cloth to dry the monitor. Donot use abrasive cleaners on the monitor. Do not use either isopropyl alcohol or solvent toclean the monitor. Use of these cleaners can cause damage to the monitors’ surface. Do notimmerse the monitor or power cord in the cleaning solution.When necessary, the monitor surfaces may be disinfected using a soft cloth saturated with a10% (1:10) solution of chlorine bleach in tap water. When all of the surfaces have beendisinfected, wipe the entire surface of the monitor using a soft cloth dampened with fresh waterto remove any trace amounts of residue and/or fumes.NOTE:Thoroughly wipe off any excess cleaning solutions. Care should be taken to prevent water orcleaning solution to run into connector openings or crevices.