CASMED 740 MONITOR SERVICE MANUAL21-02-0174 REV. 02 05/0858TEMPERATURE PROBESCAUTION:Do not soak or immerse the probe or its cable in any liquid solution. Do not attempt to sterilize.The temperature probe should periodically be cleaned by wiping with an alcohol dampenedcloth or wipe, warm water, or properly diluted non-staining disinfectant.PROBE WELL (SureTemp® Plus)CAUTION:Do not use hard or sharp objects to clean the probe well. This could damage the probe well andcause the unit to not function properly.Do not use steam, heat or gas sterilization on the probe well. Do not autoclave the probe well.As necessary, clean the inner and outer surfaces of the probe well by swabbing the surface witha cloth dampened with a mild detergent solution or a 70% isopropyl alcohol, or a 10% chlorinebleach solution, or a non-staining disinfectant.NOTE:Thoroughly dry all surfaces before re-assembling the monitor.PNEUMATIC PRESSURE CHECKA check of the monitor’s pneumatic pressure system should be performed every six (6) months.