CASMED 740 MONITOR SERVICE MANUAL21-02-0174 REV. 02 05/0886HYPOT (Monitor)NOTE:CAS Medical Systems recommends a Hypot Electrical Safety Checks be performed each timethe monitor’s case is opened or patient isolation is in question.WARNING:Do not touch the monitor when performing this test.1) Disconnect all accessories from the monitor.2) Plug the AC power cord from the CASMED 740 Monitor into the Hypot Tester.WARNING:Do not turn the monitor “ON” during a hypot test.3) Perform a Hypot Test, per manufacturer’s instructions for one (1) minute, at 2.5 KV.4) Verify that there was no arching observed.HYPOT (SpO2)WARNING:Do not touch the monitor when performing this test.1) Connect the appropriate SpO2 probe and/or cable to the monitor.2) Plug the AC power cord from the CASMED 740 Monitor into the Hypot Tester.3) Connect a lead wire from the Ground terminal of the Hypot tester to the SpO2 probe.WARNING:Do not turn the monitor “ON” during a hypot test.4) Perform a Hypot Test, per manufacturer’s instructions for one (1) second, at 1.8 KV.5) Verify that there was no arching observed.Disconnect the monitor from the test equipment.This concludes the testing to the CASMED 740 Monitor.