INSTALLATION, CONT.Load Cell Cable Connection UwithoutU RFI SuppressionThe following instructions describe the load cell connection without RFI Suppression should itnot be required.1. Remove the 14 acorn nuts securing the rear panel to the main housing.2. Referring to Figure No. 2, choose a gland connector for the load cell cable and loosen it.3. Slip the single cable from the load cell or load cell junction box through the gland connectorand into the enclosure.4. Referring to Figure No. 4, remove 3 inches (not 5 to 6 inches) of the outer insulation jacket5. Next, remove 1/4 inch of insulation from each of the 4 wires and shield (without senseleads) or 6 wires and shield (with sense leads).6. Referring to the table below (or on the circuit board) for terminal connections, connect eachof the wires to terminal block P13. Refer to Figure No. 11 for terminal block location.LOAD CELL TERMINAL – (P13)UPIN NO.U UFunctionU UPIN NO.U UFunctionU1 + EXCITATION 5 - SIGNAL2 + SENSE 6 - SENSE3 + SIGNAL 7 - EXCITATION4 SHIELD7. To terminate a wire, use a small flat blade screwdriver and press down on the release barfor the terminal. Insert the wire into the terminal opening. Remove the screwdriver,allowing the release bar to return to its original position, locking the wire in place. SeeFigure No. 5.8. Repeat the procedure until all wires are in place.9. The load cell cable shield wire should be connected to terminal 4 on terminal block P13.Load Cell Connections with Over 30 Feet of CableFor installations with over 30 feet of cable between the indicator and the load cells, sensewires should be used. The sense wires must be connected between the +SENS, -SENSterminals on the indicator and the +EXCITATION, -EXCITATION wires of the load cells or the+SENS, -SENS terminals of the load cell trim board or the section seal trim board. For theindicator to use the sense wires, the +SENS jumper J7 and the -SENS jumper J9 must beopen (see Figure No. 11).8200-M538-O1 y 225 Installation Page 12