8200-M698-O1 Rev B 225 Installation and Technical94PRESET WEIGHT COMPARATORS SETUPMode of Operation = 3 (Preset Weight Comp)With SETUP MENU #1 displayed the current setting for the 10. MODE OF OP= parameterwill be shown. Press the 1 and 0 keys and then the ENTER key. The display will change toshow the Mode Of Operation Menu. Press the 3 key and then the ENTER key toselect Preset Weight Comp. The display will change to show the Preset WeightComparator Menu below.Preset Weight Comparator Menu1. OUTPUTS=XX2. BAL PRNT=XXXEnter Selection: 0 ^ EXIT1. OUTPUTS (NUMBER OF PRESET WEIGHT COMPARATORS “PWC”)With the Preset Weight Comparator Menu displayed the current setting for theOUTPUTS= parameter will be shown. Note that XX is the current value. If thesetting displayed is acceptable, proceed to the next setup parameter.Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enterthe value for the number of PWC outputs and then press the ENTER key to save thenew setting and return to the Preset Weight Comparator Menu. Allowable values are1 through 16.The output state of each PWC before reaching the preset weight is set in the DIO SETUPMENU. Refer to the DIO SETUP (PWC Status) section of the CALIBRATION AND SETUPsection of this manual.The relay state is based on using Cardinal Scale relays (p/n 6850-1013). If usingrelays other than those supplied by Cardinal Scale, refer to the relay manufacturersspecifications.NOTE: When power to the 225 is lost, the output returns to a Low State condition.