8200-M698-O1 Rev B 225 Installation and Technical80CALIBRATION AND SETUP, CONT.Dual-Point with Zero (First Zero) CalibrationCAL 1= (FIRST CALIBRATION WEIGHT)With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then theENTER key. The display will change to show the settings for capacity, interval, decimalpoint position and CAL 1=0.0. This is the first of two calibration weights. It could beZERO (no load) or the TEST load (weight).1. If the first calibration weight is to be ZERO (no load), press the ENTER key. The displaywill flash CALIBRATING... for a few seconds and then change to show CAL 2=.2. If the first calibration weight is to be the TEST load (weight), place the weights on thescale.3. Using the numeric keys, enter the value of the TEST load (weight) and then press theENTER key.4. The display will flash CALIBRATING... for a few seconds and then change toshow CAL 2=.CAL 2= (SECOND CALIBRATION WEIGHT)The display will show CAL 2=XXXXXXX. This is the second of two calibration weights.It could be ZERO (no load) or the TEST load (weight).1. If the second calibration weight is to be ZERO (no load), press the 0 key and then theENTER key. The display will flash CALIBRATING... for a few seconds and thenchange to show CAL 3=.2. If the second calibration weight is to be the TEST load (weight), place the weights on thescale.3. Using the numeric keys, enter the value of the test load (weight) and then press theENTER key.4. The display will flash CALIBRATING... for a few seconds and then change toshow CAL 3=.CAL 3= (LAST CALIBRATION WEIGHT)The display will show CAL 3=. This weight is not used. Press the ENTER key to skipCAL 3=. The display will change to show the SCALE 1 CALIBRATION MENU. NOTE: IfSpan Adjustment is required, proceed to the FINE SPAN ADJUSTMENT on page 70.