8200-M698-O1 Rev B 225 Installation and Technical76CALIBRATION AND SETUP, CONT.CALIBRATEWith the SETUP MENU #1 displayed, press NEXT (Navigation Keys Down Arrow) toproceed to SETUP MENU #2.SETUP MENU #21. SERIAL2. DIO3. PRINT TABS4. SETUP SCALE 15. CALIBRATE 1Enter Selection: 0 ^ PREV ` EXIT5. CALIBRATE 1With the SETUP MENU #2 displayed, press the 5 key and then the ENTER key to proceedto SCALE 1 CALIBRATION MENU.SCALE 1 CALIBRATION MENU1. Swt= XXXXXXX 6. C1=XXX2. Sct= XXXXXXX 7. C2=XXX3. Zct= XXXXXXX 8. C3=XXX4. FINE SPAN 9. C4=XXX5. HI RES 10. CALIBRATEEnter Selection: 0 ^ EXIT1. Swt=XXXXXXX (SPAN WEIGHT)2. Sct=XXXXXXX (SPAN COUNT)3. Zct=XXXXXXX (ZERO COUNT)The calibration parameters (Swt, Sct and Zct) represent the test load (weight) and theanalog to digital readings at zero and test load. These values are established when thescale is calibrated. By recording the parameter values, the indicator can be returned to itspresent calibration settings without using test weights simply by entering the values.6. C1=XXX7. C2=XXX8. C3=XXX9. C4=XXXThe calibration “C” numbers (C1, C2, C3 and C4) represent the present calibrationsettings. Each number may be up to three digits in length. The “C” numbers areestablished when the scale is calibrated. By recording these numbers, the indicator can bereturned to its present calibration settings without using test weights simply by entering the“C” numbers.CALIBRATION “C” NUMBERSCALIBRATION PARAMETERS