8200-M698-O1 Rev B 225 Installation and Technical110PRESET WEIGHT COMPARATORS OPERATIONThe Model 225 indicator has sixteen (16) outputs, which can be configured during the setup ofthe weight indicator to perform as Preset Weight Comparators “PWC”. If the PWC feature wasselected during setup, the indicator will compare each enabled preset weight value with thedisplayed weight and then output a signal for each preset based on the comparison results.Enter Preset ValuesWith the indicator in normal operations mode, press the PRESET key.1.P=XXXXXX 7.P=XXXXXX 13.P=XXXXXX2.P=XXXXXX 8.P=XXXXXX 14.P=XXXXXX3.P=XXXXXX 9.P=XXXXXX 15.P=XXXXXX4.P=XXXXXX 10.P=XXXXXX 16.P=XXXXXX5.P=XXXXXX 11.P=XXXXXX 17. SET TRIM6.P=XXXXXX 12.P=XXXXXX 18. SET SCALEENTER PWC# TO CHANGE: 0PRESET TRIM SCALE EXIT1. P=XXXXXX (PRESET 1 WEIGHT VALUE)through16. P=XXXXXX (PRESET 16 WEIGHT VALUE)1. The display will change to show the values for the enabled PWCs. Note that XXXXXXis the current value for each preset.NOTE: In the example shown above, all 16 preset values are shown enabled.2. If the values displayed for each preset are acceptable, press the ENTER key to exit andreturn to normal operation.3. Otherwise, to change a preset weight value, use the numeric keys to select the PWC#(1 to 16) to change and then press the ENTER key. The display will change to showthe current value (PWC #= XXXXXX) of the preset selected.4. Using the numeric keys, enter the new value for the preset and then press the ENTERkey to save the new setting.5. Continue with this procedure until a value for each preset to be used has been entered.6. After the last preset value has been entered, press the TRIM soft key to enter the presettrim weight values or press the ENTER key to exit and return to normal operation.NOTE: Enter a zero (0) then press the ENTER key if the preset weight comparator isnot to be used.