ES601-US OPERATING MANUALAutomated Testing/Chapter 6 # Page 106Figure 72 - ECG AutoSequence in progress.Following the optional physical inspection, the ECG test sequence begins. At each step of the test sequence,the display of ES601-US will indicate the waveform and output amplitude or "scale" of the ECG simulator.The displayed amplitude corresponds to peak QRS on Lead II of the ECG.Throughout the ECG AutoSequence, PASS, FAIL and FAULT FOUND AND CORRECTED buttonsare available. Pressing the PASS button indicates the DUT responds correctly to the waveform output bythe ECG simulator. FAIL indicates the waveform is not correctly displayed on the DUT. Pressing theFAULT FOUND AND CORRECTED button means that a problem seen with the DUT was resolved bythe user (for example, a bad lead may have been replaced).Upon completion of the ECG AutoSequence, operation of ES601-US will depending on settings configuredvia the SYSTEM SETTINGS menu. After a test sequence has ended the user may be optionally promptedto input concluding comments about the ECG test as well as a labor time ("user time"). After saving theTest Record, the user may be prompted to print a PM tag or label, which can then be applied to theequipment tested. Refer to Section 6.1 AutoSequence Overview for further details.