ES601-US OPERATING MANUALES601pc-US Software/Chapter 8 # Page 163Figure 114 - ECG AutoSequence Editor.8.8 ECG AutoSequence EditorIn addition to an electrical safety analyzer, ES601-US incorporates an electrocardiogram (ECG) simulator.The ECG Simulator of ES601-US generates 40 different test signals the user can activate manually using amenu system. In addition to the waveform, the amplitude of the output signal can be set to one of threevalues: 1 mV (standard), 2 mV, or 0.5 mV (in Lead II).As an alternative to manually operating the simulator, the user can run an ECG AutoSequence in whichES601-US can be stepped through a predetermined list of up to 24 selected waveforms of given amplitudes.At each step of the ECG AutoSequence, the user can enter either a "PASS" or "FAIL”. At the end of theECG AutoSequence, the user can optionally enter a comment of up to 40 characters (e.g. "LEAD III NOTWORKING").ES601-US provides 6 programmable ECG AutoSequences, each having from one to 24 steps. ES601pc-USis used to select the waveforms and amplitudes for each step of each ECG AutoSequence and program thissetup into ES601-US.Upon selecting ECG AutoSequences on the main window of ES601pc-US, you are taken to the editorscreen (Figure 112).