ES601-US OPERATING MANUALES601pc-US Software/Chapter 8 # Page 171Figure 121 - Editing a data entry ("E") Checkitem.Figure 122 - Editing a remote control ("R") Checkitem.8.9.4.3 Entering the Command/PromptOp Codes P, E, K and I will allow editing or entering aPrompt. The Prompt is the text which ES601-USwill display when this step of the Autosequence is encountered (seeAutomated Testing Principles inChapter 6). In principle, the Prompt is intended to ask the technician to enter "OK", "PASS", "FAIL"comments or results such as “RF LEAKAGE IS 248 UA” and so on, which will then be saved in the TestRecord.Op Codes C, M, D and R will, in addition to the Prompt, allow editing or entering aCommand. TheCommand is used to transmit a message via a TESTER PORT to an external testing device, for example apatient simulator such as Datrend AMPS-1, to start or stop a function, send measurement data, etc. Formore information, refer to the Checkitem descriptions given inAutomated Testing Principles of Chapter6.