5 Care, cleaning and maintenance by the user5.5 Components of the supply moduleDentsply SironaOperating Instructions Simulation Unit Sim Intego120 66 09 437 D3649D3649.201.01.04.02 06.20195.5.4.2 Air jet pump with drain tankRemoving and emptying the drain tank and collecting tank of the airjet pumpAfter cleaning the suction system, the collecting tank should also beemptied on a weekly basis.!WARNINGWear gloves when performing the following work.1. Detach the drain hose from the drain tank.2. Empty the drain tank into a container with sufficient capacity.BAC3. Remove side panel A, loosen the connection for the protectiveground wire B, and remove the drain hose C from the side panel.