Maintenance3. Connect the cables and harnesses, identified in the following photograph,to the PWA.4. Position the power supply/KISS pump base on the chassis.5. Secure the power supply/KISS pump base to the chassis using its fasteners, asidentified in the following photograph.6. If it was installed, replace the KISS pump and tubing.See step 1 through step 4 in “Replacing the KISS Pump Assembly” on page 127for instructions.7. Replace the power supply assembly.See step 1 through step 4 in “Replacing the Power Supply Assembly” on page92 for instructions.8. Reattach the top cover assembly.See “Reattaching the Top Cover Assembly to the Device” on page 81 forinstructions.9. Reinsert the battery.See step 6 through step 8 in “Replacing the Battery” on page 74 for instructions.10. Continue with the appropriate functional checkout procedure for this FRU.See “Functional Checkout” on page 129 for more information.88 MAC™ 2000 ECG Analysis System 2053535-003C