MaintenanceRemoving the Power Inlet ModuleUse the following procedure to remove the power inlet module.1. Before beginning any steps, follow the instructions in “Preparing the System forFRU Replacement” on page 73.2. Remove the top cover assembly, if it is has not been removed already.See “Removing the Top Cover Assembly” on page 75 for more information.3. Detach the chassis.See “Detaching the Chassis” on page 82 for instructions.4. Remove the two M3x8 screws securing the power inlet module to the bottomcover, as shown in the following photograph.5. Remove the power inlet module.Replacing the Power Inlet ModuleUse the following procedure to replace the power inlet module.1. Position the power inlet module in the bottom cover.2. Secure the power inlet module to the bottom cover using two M3x8 screws, asshown in the following photograph.3. Reattach the chassis.See step 1 through step 6 in “Attaching the Chassis” on page 85 for instructions.4. Reattach the top cover assembly.See “Reattaching the Top Cover Assembly to the Device” on page 81 forinstructions.5. Reinsert the battery.See step 6 through step 8 in “Replacing the Battery” on page 74 for instructions.6. Continue with the appropriate functional checkout procedure for this FRU.See “Functional Checkout” on page 129 for more information.2053535-003C MAC™ 2000 ECG Analysis System 93