ABT-BA-e-13125614.3 Density determinationDensity determination with the help of the flush-mounted platform facility is describedin the following.Density determination becomes even easier when an optional set for density deter-mination is applied. For further information please refer to the operating instructionsenclosed with the set for density determination.1. Remove the below-weigh hook cover on the floor of the balance, after remov-ing the two fixing screws.2. Hang the user-prepared hanging pan from the below-weigh hook and sink thathanging pan into the sample liquid in the tank.14.3.1 Density determination of solidsCondition: Function ",d" (density of solids) is activated. See Chapter 11.3.Press the [CAL] key repeatedly until "SettinG" appears.Press the [TARE] key.Repeatedly actuate the [CAL] key until "LSG SEt" appears(Example)Press the [TARE] key. The currently set density for the liquid to bemeasured appears. In the upper part of the display panel, thesymbol and the # symbol appear in order to indicate numer-ical input status. The leftmost digit blinks.(Example)Enter density for your liquid to be measured. When the [UNIT] keyis pressed, the numerical of the blinking digit increases by 1 at atime. You can determine the value of the flashing digit, or shift theflashing digit by one position to the right, by pressing the [PRINT]key. Confirm your setting by pressing the [TARE] key.Repeatedly press the [ON/OFF] key until the balance is in weigh-ing mode.