ABT-BA-e-1312 6515.3 Remote control instructionsAttention:Inputting characters and command codes not shown here into the balance may notonly alter the previous settings but may also impair proper measurement.If by mistake characters or commands not shown here are entered into the balance,immediately unplug the power supply cable and wait about ten seconds before plug-ging again.Command Code Function DescriptionD01 Continuous output The balance continuously outputs every 110ms .D05 1 time output Corresponds to [PRINT] keyD06 Auto Print See 14.4D07 1 time output with stabilityinformationThe status of the stability mark is appended to the headof the data with output.S: when the standstill display appearsU: when standstill display does not appearD08 1 time output at stability After command input, the data are output at the firstappearance of the standstill display.D09 Halt output Auto Print and continuous output haltedQ ON/OFF switching Switches between standby status and measurementstatus.T Taring Corresponds to the [O/T] keyTS Taring after stability wait After command input, taring is done at the first appear-ance of the stability mark.C18 Measurement span calibration+ Measurements in add-up mode See ch. 14.5R Total reset All application measurements terminated and resetmg mg unitPERCENT Percent determinationPCS Parts countingCT ct unitSDENCE Density of solidsLDENCE Density of liquids% 100% settingG g, % switching- g g unit removal- mg mg unit removal- PERCENT Deleting percentage determina-tion t- PCS Piece counting removal- CT ct unit removal- SDENCE Solid specific gravity removal- LDENCE Liquid specific gravity removal