ABT-BA-e-131262Printout:---ADDON Mode---CMP001=5,0003 gCMP002=5,0017 gCMP003=5,0010 gTOTAL=15,0030 gNote: Only one function at a time may be activated (on) when working with func-tions auto print, add-on and recipe.Whilst the function is active, automatic adjustment (PSC/Clock-CAL) doesnot take place. A flashing symbol, however, indicates that adjustment isrequired.14.6 Recipe modeThe formula function allows to add on various components of a mixture. The totalweight of all components can be invoked to check (separate memory for the weightof tare vessel and recipe components).Activate function "FormU: on“ (see chap. 14.5 Add up):Whilst the recipe function is active, the display keeps showing the "FormU -on" sym-bol [ M].Operation DisplayWhen using a container, place this on the weighing plateand tare. 0.0000 g�Weigh first component. 0.5361 gWhen standstill control ( ) is complete, the weighingvalue of the first component (CMP001) is issued to theoptional printer by pressing the [PRINT] key. The dis-played value is added into the total adding memory. Af-terwards automatic taring will take place.0.5361 g0.0000 gWeigh second component 0.5422 gWhen standstill control ( ) is complete, the weighingvalue of the second component (CMP002) is issued tothe optional printer by pressing the [PRINT] key. Thedisplayed value is added into the total adding memory.Afterwards automatic taring will take place.0.54221 g0.0000 g