Keysight E8402A, E8404A C-Size Mainframes User and Service Guide 117STATus Subsystem Programming the Enhanced MonitorSTATus:QUEStionable:BLOWer:ENABle <mask>STATus:QUEStionable:BLOWer:ENABle?The STATus:QUEStionable:BLOWer:ENABle command sets the value of the enableregister for the BLOWer status group.The STATus:QUEStionable:BLOWer:ENABle query returns an integer representingthe value of the enable register for the BLOWer status group.The enable register specifies which bits in the event register can generate asummary bit. The instrument logically ANDs corresponding bits in the event andenable registers, and ORs all the resulting bits to obtain a summary bit. Summarybits are, in turn, recorded in another register. The blower summary bit is recordedin the Questionable Condition Register. Enable registers are read-write. Enableregisters are not affected by *CLS (Clear status). Querying enable registers doesnot affect them.ParametersReturned DataComments– Refer to STATus:QUEStionable:BLOWer:CONDition? for a description ofeach bit in the BLOWer status group.– If *PSC was set to 0 prior to power-down, then at power-on, the enableregisters are loaded from non-volatile memory. Therefore, any values thatyou have stored (SYSTem:NVSave command) are returned to the registers.However, if *PSC was set to 1, then at power-on the factory defaults arereturned.Name Type Range Default Description<mask> int16 0 - 32767 7(see 2ndcomment)Bit mask indicating which enable bits are set(1) or cleared (0) for the BLOWer statusgroup.Type Range Descriptionint16 0 - 32767 mask value.