190 Keysight E8402A, E8404A C-Size Mainframes User and Service GuideProgramming the Enhanced Monitor SYSTem SubsystemSYSTem:NVSaveSYSTem:NVSave causes the enhanced monitor to store many of its parametersto non-volatile memory. This includes temperature limits, current limits, totalpower limits, status enable settings, PSC flag, beep frequency, beep duration,system name, mainframe serial number, and all RS-232 settings.Comments– Non-volatile memory has a finite access lifetime. For this reason, thesevalues are saved only when explicitly requested with this command and allat one time. The commands/registers affected are:Error ConditionsThe following table lists the most common error conditions and causes. Errornumbers and corresponding messages can be found using SYSTem:ERRor?query.Register RegisterSTATus:OPERation:ENABle SYST:BEEP:FREQSTATus:QUEStionable:ENABle SYST:BEEP:TIMSTAT:QUES:BLOWer:ENABle SYST:NAMESTAT:QUES:CURRent:ENABle SYST:SNUMSTAT:QUES:TEMPerature:ENABle HIST:UNIT:TIMESTAT:QUES:VOLT:ENABle *PSCSTAT:QUES:VOLT:PTR *ESESTAT:QUES:CURR:LIM *SRESTAT:QUES:TEMP:LIM SYST:COMM:SER:REC:BAUDSTAT:QUES:POW:LIM SYST:COMM:SER:REC:BITSSYST:COMM:SER:CONT:RTS SYST:COMM:SER:REC:PACESYST:COMM:SER:ECHO SYST:COMM:SER:PARitySYST:COMM:SER:ERES SYST:COMM:SER:SBITSYST:COMM:SER:LBUFNumber Message Probable Cause-221 “Settings Conflict” The mainframe’s power is off.-311 “Memory Error” Non-volatile memory is inaccessible.