146 Keysight E8402A, E8404A C-Size Mainframes User and Service GuideProgramming the Enhanced Monitor STATus SubsystemSTATus:QUEStionable:UMCounter:TINTerval <time>STATus:QUEStionable:UMCounter:TINTerval?The STATus:QUEStionable:UMCounter:TINTerval command sets the time intervalvalue of the user maintenance counter. The user maintenance counter is acountdown timer that counts hours down to zero. The time interval is the valuefrom which the UMCounter will start. This value is stored in non-volatile memoryimmediately.The STATus:QUEStionable:UMCounter:TINTerval? command returns an integerrepresenting the time interval value of the user maintenance counter.ParametersReturned DataComment– The value 65535 means that the counter is disabled. It is shipped from thefactory disabled. Set the time interval (STAT:QUES:UMC:TINT) and thenreset the counter (STAT:QUES:UMC:TRES) to enable the user maintenancecounter. To disable the counter, send the commandSTATus:QUEStionable:UMCounter:TINTerval DEF.– The MINimum value is 1, MAXimum value is 65534, DEFault value is65535.Reset Condition*RST does not affect the UM Counter.Related CommandsSTATus:QUEStionable:UMCounter:TRESet,STATus:QUEStionable:UMCounter:TREMaining?Name Type Range Default Description<time> int16 1 to65535,MIN,MAX, DEF65535 number of hours the UMCounter will count downfrom.Type Range Descriptionint16 User Maintenance Counter