Keysight M8070A Programming Guide 217SCPI Command Reference 5:OUTPut:DEEMphasis:PRECursor(*)[?]Syntax :OUTPut:DEEMphasis:PRECursor(*) ‘identifier’, :OUTPut:DEEMphasis:PRECursor(*)? ‘identifier’InputParameters‘identifier’: 'M*.DataOut1' or 'M*.DataOut2': Set selected pre-cursor value.Return Range PRECursor1: -12.04 dB to 12.04 dB | 25% to 400%PRECursor2: -6.02 dB to 6.02 dB | 50% to 200%Description This command programs the pre-cursor value of the currently selectedpre-cursor (pre-cursor 1 or pre-cursor 2).This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.For M8045A, the pre-cursor values cannot be changed. You can set thepre-cursor values by adjusting the coefficient values.Example :OUTP:DEEM:PREC2 ‘M1.DataOut2’, 3.1:OUTPut:DEEMphasis:POSTcursor(*)[?]Syntax :OUTPut:DEEMphasis:POSTcursor(*) ‘identifier’, :OUTPut:DEEMphasis:POSTcursor(*)? ‘identifier’InputParameters‘identifier’: 'M*.DataOut1' or 'M*.DataOut2': Set selected post-cursor value.Return Range POSTcursor1: -20 dB to 20 dB | 10% to 1000%POSTcursor2, POSTcursor3: -12.04 dB to 12.04 dB | 25% to 400%POSTcursor4, POSTcursor5: -6.02 dB to 6.02 dB | 50% to 200%Description This command programs the post-cursor value of the currently selectedpost-cursor (post-cursor 1 through post-cursor 5).This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.For M8045A, the post-cursor values cannot be changed. You can set thepost-cursor values by adjusting the coefficient values.Example :OUTP:DEEM:POST4 ‘M1.DataOut2’, 2.3