Click Mode Operating Instructions 10-910.5.1 Stop and Pause If Stop has been selected during a measurement, the test can then beaborted or it is possible to go immediately to the next step.If Paused, the test can be resumed at any convenient time.10.6 Report After a Click test the PMM 9010 reports all the relevant data.Due to the complexity of the test and to the various conditions that mayhappen during the measurement, several reports may be issued.Here following some examples well representing the completeness and theaccuracy of the analyzer.10.6.1 Fail duringdetermination ofthe click rate NIn case a fail is recorded during the determination of the click rate N at thefrequency of 150 kHz, a report like the following is displayed (which is justan example with a not-realistic N number):If the fail occurs during the determination of the click rate at 500 kHz, thereport may be like this:10.6.2 Report after asuccessful test withless than 5instantaneousswitching at onefrequencyIf the determination of the click rate N shows that at a given frequency (500kHz in this examples) there are less than 5 instantaneous switchingoperations, there is no need of any further test at that frequency orfrequencies (the investigation frequencies are 150 kHz for the interval from150 to 500 kHz, and 500 kHz for the interval from 500 kHz to 30 MHz).In the example, L q is calculated for 150 kHz and no tests are done at theother frequencies, as the number in instantaneous switching at 500 kHzduring the observation time is less than 5.To complete the test of the example above, only the steps 1, 2 and 3 wereperformed.