A-C-8 Four Channels Click Meter OptionA-C.5 Operation As soon as the Setup has been done it is possible to start the test.First of all the PMM 9010 will determine the click rate N, and – if all theconditions apply – after that the analyzer will immediately continue measuringthe number of clicks exceeding the upper quartile limit L q.Simply select Start: the PMM 9010 will automatically take care of all thenecessary steps, reporting at the end of the test all the relevant results.In this way the PMM 9010 does the determination of the click rate at the twofrequencies of 150 kHz and 500 kHz simultaneously, and then it executes thetest.A-C.5.1 SmartMeasureThe “Smart Measure” is a PMM 9010 unique feature which allows a relevanttime saving in Click measuring approach.PMM 9010 can trace and memorize eventual saturation just got duringautomatic measurement test procedure, in a way that, in case no saturationconditions at all have been detected, the so got test result can be useddirectly to soon calculate the new limit Lq, without the need of repeating thetest for it.To get the system working properly it is necessary the PMM 9010 runs afirmware revision 1.85 or following.Be sure not to overload PMM 9010/Click4E: the input signal should notexceed the maximum level indicated in the main specifications.Also do not apply any signal to RF generator output connector.Click Meter Reports can be remotely downloaded to the PC using thePMM Emission Suite Software.A-C.6 OperatingManualChangesInstruments manufactured after the printing of the User’s manual mayhave a serial number prefix not listed on the title page; this indicatesthat instruments with different Serial Number prefix may be differentfrom those documented in the manual itself.