12ModeUnitsSetup PrintEnterModeUnitsSetup PrintEnter3.3.4 Span CalibrationSpan calibration utilizes two calibration points, one at zero and the other at full span or incremental values startingat 25 % of full capacity. As an example, an 8.1 kg balance will accept either 2 kg, 4 kg, 6 kg or 8 kg for spancalibration. Values which are below or in between will not be accepted and the balance will display its maximumcapacity. When LFT is set on, only full span calibration value can be used. Sample display illustrates an 8.1 kgbalanceStablegSetupUse keys to change, press EnterChangeSetupUse keys to change, press EnterChangeUse keys to change, press EnterChange SetupSetupUse keys to change, press EnterChangeSetuppress EnterChangegSetupChangeProcedure• Press the Setup button, CAL is displayed.• Press Enter button, CAL TYPE is displayed.• Press Enter button, CAL TYPE InCAL is displayed.• Press button to select SPAN calibration, CALTYPE SPAN is displayed.• Press Enter button, WORKING is displayed.NOTE: If a weight is left on the pan, the balance willdisplay CLR PAN (remove the weight from the pan). Thebalance automatically resumes calibration.• Display changes to PUT WT 8000 g. The displayedweight is the full capacity of the balance.• Place specified calibration mass on pan.NOTE: For an 8.1 kg balance, either 2 kg, 4 kg, 6 kg or8 kg can be used for span calibration. The PUT WEIGHTmessage indicates the calibration mass that is on thepan.• Press Enter button, WORKING is displayed. After afew seconds CAL SET is displayed, the display thenreturns to WEIGH mode.• Span calibration is completed.• Remove calibration mass from the pan.