395.3 Error Codes ListError Codes ListThe following list describes the various error codes and which can appear on the display and the suggestedremedy.Data Errors1.0 Transient error (hardware error, probably static discharge). If error persists, the balance must beserviced.1.1 Balance temperature transducer hardware error.Tare Errors2.0 Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needsrecalibration.Calibration Errors3.0 Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.RS232 Errors4.4 RS232 buffer is full.User Errors7.0 User entry out of bounds.7.2 Number outside of display capacity.Over-Under Load Errors8.0 Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balancemust be serviced.8.1 Hardware error caused by an internal weight signal which is too high. Check load on the pan which maybe excessive. If error persists, the balance must be serviced.8.2 Power-on load out of specification (LFT only)8.3 Rated capacity exceeded. Remove excessive weight from pan.8.4 Underload condition on balance. Check that the proper pan is installed.8.5 Internal calibration weight internal sensor indicated its weight on the pan.CheckSum Errors9.1 Bad factory checksum. If error persists, have the balance serviced.9.2 Bad factory checksum. If error persists, have the balance serviced.9.3 Bad factory checksum. If error persists, have the balance serviced.9.4 Factory internal calibration data failed checksum. This failure will disable access to the InCALTM feature(if installed).9.5 Factory calibration data failed checksum.9.8 User calibration data failed checksum.9.9 Factory temperature compensation data failed checksum.5.4 Information MessagesCAL NOW If InCAL TM (internal calibration) is installed. Message to recalibrate the balance. Themessage will remain until calibrated.WARM UP The user tried to perform an internal calibration and this message will be flashed in the 14segment field. The balance requires a 7 minute warmup period. During warmup the usercan not select InCALTM from the menu.SAVED This message is flashed when an item is changed in the menu and the new value iswritten to the EEPROM.LOCKED This message is flashed when an item can not be changed in the menu because themenu is locked and the Lock Switch is set locked.LOW REF The message is flashed in parts counting or percent when the calculated reference weightis very low.UNSTBLE This message is flashed when the balance was unable to aquire stable data during internalcalibration.