163.3.8 Calibration GLP PrintoutIf any option in the GLP Set Menu is turned On, GLP automatically prints data after calibration is completed.- - - - - SPAN CAL - - - - - -12/01/97 1:00:00 PMBal Id 1234Cal: 8000.00 gOld: 8000.00 gDif: 0.00 gWt. Ref......................................USER NO 2056853PROJ NO 100012Name........................................- - - - - END - - - - -- - - - - INCAL - - - - - -12/01/97 1:00:00 PMBal Id 1234Cal: 8000.00 gOld: 8000.00 gDif: 0.00 gWt. Ref......................................USER NO 2056853PROJ NO 100012Name........................................- - - - - END - - - - -InCAL TM Calibration PrintoutWhen performing InCALTM calibration with all GLP optionsturned on, a printout is automatically made after thecalibration is completed.- - - - - LIN CAL - - - - - -12/01/97 1:00:00 PMBal Id 1234Cal: 8000.00 gOld: 7999.08 gDif: 0.02 gWt. Ref......................................USER NO 2056853PROJ NO 100012Name........................................- - - - - END - - - - -- - - - - CAL TEST - - - - - -12/01/97 1:00:00 PMBal Id 1234Cal: 8000.00 gAct: 8000.02 gDif: 0.02 gWt. Ref......................................USER NO 2056853PROJ NO 100012Name........................................- - - - - END - - - - -Span Calibration PrintoutWhen performing Span calibration with all GLP optionsturned on, a printout is automatically made after thecalibration is completed.Linearity Calibration PrintoutWhen performing a Linearity calibration with all GLPoptions turned on, a printout is automatically made afterthe calibration is completed.Calibration Test PrintoutWhen performing a Calibration Test with all GLP optionsturned on, a printout is automatically made after the cali-bration is completed.