Breaking capacity min. 50 mW, max. 500 VA, max. 54 W DC1)Min. switching hysteresis 0.5 %Current outputsQuantity 1x outputRange 0/4 … 20 mAResolution 0.1 % or 20 μAMax. load 500 OhmFailure message 22 mALinearity error 0.1 % (relating to 20 mA)Temperature error 0.01 %/K (relating to 20 mA)IndicationsMeasured value indication- LC display (45x45 mm) digital and quasianalogue indication- Max. indicating range -9999 … 9999LED displays- Status indication operatingvoltage1x LED green- Status indication fault sig-nal1x LED red- Status indication operatingrelay 1/22x LED yellowOperationAdjustment elements 2x keys in the front, 1x function switchAmbient conditionsAmbient temperature -20 … +60 °C (-4 … +140 °F)Storage and transport tem-perature-40 … +80 °C (-40 … +176 °F)Electrical protective measuresProtection- Wall/rail mounting IP 20- Front panel mounting IP 40Overvoltage category IIProtection class II1) If inductive loads or stronger currents are switched through, the gold plating on the relay contact surface will bepermanently damaged. The contact is then no longer suitable for switching low-level signal circuits.30 VEGAMET 381Supplement27567-01-070216