Piranha4 2K and 4K Color Camera User's Manual 125Teledyne DALSA 03-032-20180-06Cleaning the sensor windowRecommended Equipment Glass cleaning station with microscope within clean room. Ionized air gun. For example, 3M ionized air gun. Ionized air flood system, foot operated. Cotton swab. For example, Swab Huby340 CA-003. Single drop bottle (FD-2-ESD). Optic cleaning fluid. For example, the E2 Eclipse optic cleaning system.Procedure Use localized ionized air flow directed on to the glass during sensor cleaning. Blow off mobile contamination using an ionized air gun. Place the sensor under the microscope at a magnification of 5x to determine thepresence of any remaining contamination. Clean the contamination on the sensor using one drop of cleaning fluid on aswab. Wipe the swab from left to right (or right to left but only in one direction). Dothis in an overlapping pattern, turning the swab after the first wipe and witheach subsequent wipe. Avoid swiping back and forth with the same swab inorder to ensure that particles are removed and not simply transferred to a newlocation on the sensor window. This procedure requires you to use multipleswabs. Discard the swab after both sides of the swab have been used once. Repeat until there is no visible contamination present.