Chapter 3. Initial Setup and Programming42 AT600 User’s Manual3.6.1c Calibrate the Output3.6.1d Setting Error HandlingUse the Calibrate menu to trim the analog output to your measurement system.In the Analog Output menu, select Calibrate and press []. A screen similar tothe following appears.Select 4 mA to trim the 4 mA level, 20 mA to trim the 20 mA level, orPercentage of Scale to test linearity. Select the appropriate option and press [].A screen similar to the following appears.Read the analog output by a multimeter or other device. And input the currentvalue.Use the [] or [] arrow key to select the digit to be changed, use the [] or[] arrow keys to change the Calibrate setting value and press []. Press [] toreturn to the previous display.Repeat these steps until the output value is correct.To specify the error handling status in the Analog Output A menu, select ErrorHandling and press []. The following screen appears.Selecting Low will force the Analog Output to 3.6 mA or below, while High willforce it to 21.6 mA or above. Select the appropriate status and press [].Analog Output AMeasurementBase ValueFull ValueCalibrateCalibrateCalibrate 4mACalibrate 20mAPercentage of ScaleCalibrate 4mAUNIT: mA4.000]MOVE]UNDO [ ]SAVE ]]]MODF[Analog Output ABase ValueFull ValueCalibrateError HandlingCalibrateLowHighHoldOther