AquaTrans™ AT600 User’s Manual 129Chapter 6. CommunicationCommand 184 (0xB8): Write Digital ConfigurationThis command is to write the digital output configuration.Table 80: Request Data Bytes for Write Digital ConfigurationByte Format Description0 Unsigned-8 Channel Number(1/2)1 Unsigned-8 Digital Output type:0: Off; 1: Pulse; 2: Frequency; 3: Alarm;Table 81: Response Data Bytes for Write Digital ConfigurationByte Format Description0 Unsigned-8 Channel Number(1/2)1 Unsigned-8 Digital Output type:0: Off; 1: Pulse; 2: Frequency; 3: Alarm;Table 82: Command-Specific Response Codes for Write Digital ConfigurationCode Class Description0 Success No Command-Specific Errors1 Undefined2 Error Invalid Selection3-4 Undefined5 Error Too Few Data Bytes Received6 Error Device-Specific Command Error7 Error In Write Protect Mode8-15 Undefined16 Error Access Restricted8-127 Undefined