8 - 2 Quick Check® 600/800 Series User’s GuideReflectance-Based InformationWhen connected to a QC3800V, the Main display indicates the Series 800verifier model, and the bottom line displays:Linear ImagerOnce a successful scan is performed, the Main display shows, for example:Note that the second line, where the reflectance information would normally bedisplayed, shows ---, which indicates that the reflectance information is notmeasured. The reflectance information is measured and displayed when aMouse Wand or Pen Wand is connected to the Series 800, but not when aQC3800V is used. No reflectance-based values are displayed or printed. Thisincludes Rmin, Rmax, Print Contrast Signal, Symbol Contrast, Edge ContrastMinimum, Modulation, and Defects.ScanProfile OutputNo ScanProfile Output is available in either the Printer Output modes or the"PROF to PC" mode (as used by QCViewerII software).Percent Decoded/Scanned ModeNo Reflectance Calibration is necessary when the QC3800V is connected.When a Mouse Wand or Pen Wand is reconnected, however, the reflectancecalibration is necessary.When using the QC3800V, a special Percent Decoded/Scanned Mode replacesthe Reflectometer Mode found on a Series 600 verifier. When the QC3800V isconnected, the second level (Print Results) menu changes from:* Print Results* Calculate "X"* Reflectometer* Change SettingsC39:ABCD012345678Dim --- FmtDecod’ty 75% WN Ratio 2.3 OK! PreviousNext |