DAB-BA-e-1813 156 Transport and storage6.1 Testing upon acceptanceWhen receiving the appliance, please check packaging immediately, and theappliance itself when unpacking for possible visible damage.6.2 Packaging / return transport Keep all parts of the original packaging for a possibly requiredreturn. Only use original packaging for returning. Prior to dispatch disconnect all cables and removeloose/mobile parts. Secure all parts against shifting and damage.7 Unpacking, Setup and Commissioning7.1 Installation Site, Location of UseThe instrument is designed to achieve reliable weighing results under normalconditions of use.You will work accurately and fast, if you select the right location for your moistureanalyser.On the installation site observe the following:! Remove explosion prone and easily flammable material in the immediatevicinity.Emerging vapours, sample dish and all parts of the sample chamber are hot!! Protect the instrument against direct draughts due to open windows anddoors.! Avoid extreme heat and temperature fluctuations e.g. due to installation nextto radiators.! Do not expose the instrument to extreme dampness for longer periods of time.Non-permitted condensation (condensation of air humidity on the instrument)may occur if a cold instrument is taken to a considerably warmer environment.In this case, acclimatize the disconnected instrument for ca. 2 hours at roomtemperature.! Avoid direct sunlight! The air humidity should be between 45% and 75%, non-condensing.