32 DAB-BA-e-18133. Selecting switch-off criterionA switch-off criterion defines under which conditions the device should cancel thedrying process. Switch-off criteria save continuous time controls and manualcancelling of drying processes. Furthermore they ensure that measurements arealways finished under the same conditions, thereby providing repeatablemeasurements. After acknowledgement of the selected heating profile by , the currentlyset heating profile and switch-off criterion flashes. Using arrow keys select required switch-off criterion and confirmby .Selectable settings suitable as switch-off criteria include:Automatic switch off(2mg weight loss in 45 s)This switch-off criterion is based on aweight loss per time unit. As soon as themedium weight loss drops below thedesired value per unit of time, themeasurement will be finishedautomatically.Time-controlled switch off,selectable 3 min – 99 minWhen this switch-off criterion is selectedthe measurement will continue until the setdrying time has lapsed Manual switch-off by pressing stop key PreviousNext |