DAB-BA-e-1813 2910 Define drying parametersIn chapter 9 you have already carried out a first measurement with the factorysettings.The device provides a great variety of setting options that allow you to adapt a dryingmethod to your sample.Three heating profiles (standard drying, soft drying, fast drying) and three switch-offcriteria (time-controlled, automatic, manual) can be selected.1. Enter ident number for drying method16 memory locations, which can be simply called up and started as necessary underthe saved ident number (0 –F), are available for complete drying methods. After pressing the first ident number <0> will be displayed.This memory location is provided for the factory settings (standard drying120°C, automatic switch-off) and cannot be changed. Use the arrow keys to select the desired ident number andconfirm by . The currently set heating profile flashes.