DAB-BA-e-1813 218 Adjustment8.1 Weight adjustmentThe weight adjustment of the integrated balance is not absolutely necessary for acorrect moisture determination, as this measurement is only carried out relatively.The instrument calculates the weight of the sample before and after the dryingprocess and the moisture content is determined with help of the ratio between wetweight and dry weight.The instrument however should be adjusted, if this is required due to the qualitysystem used by you.Procedure:• Observe stable environmental conditions. A warming up time (see chapter1) is required for stabilization.• Carry out adjustment with placed sample dish. Ensure that no objects arewithin the sample dish.• Carry out adjustment as near as possible to the balance’s maximumweight (recommended adjustment weight see chap. 1). Info about testweights can be found on the Internet at: http://www.kern-sohn.com. Press and keep pressed until „CAL“ is displayed. Wait until the weighed value for the required adjustment weight appearsflashing.Sample indication DAB 100-3