26 DAB-BA-e-1813Begin temperature calibration: Switch on temperature calibration set. Close heating hood and press . The instrument will be automaticallyheated to the set temperature. The current temperature and the elapsed timewill be indicated in the display. After approx. 15 minutes finish by . Compare the temperature value ofthe moisture analyser with that of the temperature calibration set. At adifference of ±5°C we recommend to adjust the temperature, see followingchapter.8.2.2 Temperature adjustmentIf the admissible deviation is exceeded / not reached during temperature calibrationthe temperature of the instrument can be adjusted as described below.1. Switch off instrument.2. Press , quickly release, then press and at the same time.„8.2X“ will be displayed. If not, separate device from power supply and restartstep two.3. Adjust temperature using arrow keys . Each time you press thebutton the temperature will be increased or reduced by 1°C.4. Confirm input by .