7 - 1LHM2A*RDA001A/E7.1 Preparation for Setting and Adjustment1. Prepare necessary adjustment and measurement jigs (see 8.5).2. Connection with the Lex 10001) Connect the provided RS-232C cable (41268-E011) between the Edger port of theLex Drill and the Drill port of the Lex 1000.3. Communication settings in the Lex Drill1) Turn off the Lex Drill.2) Set the DIP SW 1-1 on the main board(44101-BA01) to the LOW position (upposition).4. Communication settings in the Lex 10001) Turn on the Lex 1000.2) Press to display the Menu screen.3) Press to select “CommunicationSetting”, then press .㪩㪪㪄㪉㪊㪉㪚㩷㪚㪘㪙㪣㪜䋨㪋㪈㪉㪍㪏㪄㪜㪇㪈㪈㪀59 .19Main board (44101-BA01)§ 7 ADJUSTMENT