445.0 SetupSUMMARY: The Setup Menu consists of the following items; the following sections providefurther explanation:Scale: The following electronic scale is recommended for the FilterTec Plus and will ship withmost systems: Ohaus Adventurer Pro. Proper communications are set by selecting Ohaus 3as the Scale Manuf. Other scales, Mettler, Ohaus or Sartorius may be used. Submenu allowssetting of Scale Manuf., Units, Alarm, and Tare. Default = “Ohaus3”.Clock: Set the time of day (military), day, month, and year. Print Enable allows choice of Timeof Day, or Relative (Run) Time for printout and display. In most cases the clock will be set at thefactory for the destination time zone. Default = Time of Day.System Test: Allows testing of the I/O’s of the FilterTec Plus, requires purchase of IQ/OQDocument. Use Test Mode if needed for trouble shooting.Test Mode: Allows independent testing of the FilterTec Plus I/O’s. Motor, Keypad, Scale,Pressure, Temperature, Valves, Analog, TTL switches.Ethernet: Allows setting of the IP Address, Subnet Mask and Gateway values for ModbusTCP/IP communication via the Ethernet Port.Printer: Select communications parameters for Parker printer (P/N 080-095) or PC. Defaultsettings are Print Time (Default = 10 sec.), Type (Seiko), Baud Rate (9600), Stop Bits (2),Parity (None), Word Length (8), Print Delay (0 sec).Analog: Allows setting of Hi and Lo Range, as well as Hi and Lo Limits, Alarms and Zeroing ofall three available 4-20 ma analog inputs.Temperature: Select an Offset for the SciTemp Temperature sensor if needed.Press. Sensor: Allows user to Zero all three SciPres pressure sensors. (Span is used forfactory calibration.) Source is used to select the desired sensor for related alarms and control inall modes. Units: Choose from Psi (default), Bar, or Kpa. Range: Default is 60, can be setlower, will require re-calibration of the input. Contact Parker Customer Service for assistance..Pump: Select the following user preferences: Keypad Beep: (On/Off), Switch Configuration:(Level / Pulse), Switch Polarity: (Normal/Inverted), TTL1:On-Off: (Yes/No), Set Yes ifcontrolling another pump as a slave, set NO if controlling Rotary Selector Valve “W”. MotorStart: (Hard / Soft-Ramp), Motor RPM: (3400, 600, 160, 8), Pump Head:(Peristaltic/RH1/RH0/RH00), Pump Tubing: (13/14/16/25/17/18/15/24/35), Power Up:(Mode/Menu/Run), External Run-Stop: (Pulse/Level), ASCII Feedback (On / Off), FactoryReset: (Resets all variable parameters to their original factory defaults).Scale2 and Scale3: The following electronic scale is recommended for the FilterTec Plus andwill ship with most systems: Ohaus Adventurer Pro. Proper communications are set byselecting Ohaus 3 as the Scale Manuf. Other scales, Mettler, Ohaus or Sartorius may be used.Default = Ohaus3 for both scales. Set to “None” if using fewer than three scales, or a “ScaleError” will occur at initialization.