525.7 Setup: Pressure SensorPressure Sensor is used to configure settings related to the SciPres Disposable PressureSensors. The following can be configured; Range, Units, Source (control and alarms), Filter(signal noise) and Zero/Span. Range: Set the overall pressure range, the Default is 60.0 psi, and is the maximum.Changing this setting requires re-calibration of the sensor signals. Units: Select between Psi, Bar, and KPA. Default = Psi. Source: Controls the source pressure sensor for control in Constant Pressure Mode, andfor all pressure alarms. Select between P1, P2 or P3. Default = P1 Zero: Used to “Zero” the sensors to establish the zero offset cause by the circuitry and anyinherent hydraulic pressure in your tubing beyond the pump head. Connect the sensor, andremove any pressure from the system. Select the sensor, P1, P2 or P3 and press “Zero”.You are prompted to confirm the action. “Span” should not be used on a routine basis, as itsets the max input value. NOTE: Do not Zero P1, P2 or P3 unless a sensor is connected.The display should read “--.-“ with no sensor connected. If it reads 0.0 in this state, connectthe sensor and re-zero it. Filter: This feature filters out the pulsations in the pressure signal due to the peristalticnature of the pump head. This provides better control and easier to read values, aswell as improved graphed data. Select from 1 to 7, 1 = no signal filter, 7 = maximum filter.Default = 4.- SETUP-PRESS.SENSORUP DOWN SELECTPRESS RANGE: 60.0UP DOWN SELECTPRESSURE: RANGEUP DOWN SELECTPRESSURE UNITS: PSIUP DOWN SELECTPRESSURE: UNITSUP DOWN SELECTPRESS SOURCE: P1UP DOWN SELECTPRESSURE: SOURCEUP DOWN SELECTPRESSURE: ZEROP1 P2 P3PRESSURE: ZEROUP DOWN SELECTPRESS FILTER: 4UP DOWN SELECTPRESSURE: FILTERUP DOWN SELECTPRESSURE: 1 0.0ZERO SPAN EXITSELECTDOWNDOWNDOWNDOWNP1