INTRODUCTIONCarl Zeiss Equipment safety notices Axio Examiner10 M60-2-0003 e 05/2012Gas discharge lamps as, e.g., HBO 100 emit ultraviolet radiation which might cause burning tothe eyes and to the skin. Accordingly, do not look directly into the light source, and avoid anydirect incidence of light on your skin. Whenever working on the microscope, always use theprotection devices belonging to the equipment (e.g., specific attenuating filters or thefluorescence protection screen). Gas discharge lamps are under a high internal pressure whenhot. Accordingly, they should only be replaced after having cooled down, and using bothprotective gloves and a protective mask.When using a fluorescence filter, do not remove the heat filter for protection against the heatradiation emitted by the microscope lamp since fluorescence filters are heat sensitive, and heatmay impair proper functioning.Avoid touching the hot lamp housing. Prior to lamp replacement, unplug the power plug fromthe corresponding power supply unit, and allow approximately 15 minutes for the lamp to cooldown.Dust and contamination may impair the equipment's operational reliability. So, the equipmentshould be protected against any such influence as far as possible. Cover the equipment withthe dust cover whenever it is not in use. Before covering the equipment, always check that alllighting devices have been switched off.Clogged or covered venting slots may lead to heat buildup which may damage the equipmentand even cause a fire in extreme cases. Venting slots should always be kept free and open, andno object should be inserted or allowed to fall in.This equipment must not be operated unless by duly instructed personnel. Such personnel mustbe instructed about possible hazards related to microscopy and to the corresponding field ofapplication. Axio Examiner microscopes are precision instruments, and their operability may beimpaired, or the equipment may be destroyed by any inexpert treatment or intervention.Make certain to read the safety data sheets for Immersol 518 N® and 518 F®.Immersol 518 N® and 518 F ® immersion oils are irritating to the skin. Avoid any contact withskin, eyes, or clothing.Wash with abundant water and soap in the event of any skin contact.After eye contact, immediately rinse with abundant water for not less than 5 minutes. Consulta specialist physician if irritation persists.Appropriate disposal (Immersol 518 N ® and 518 F®):Do not allow to be discharged into surface waters or sewage drains.Defective microscopes do not belong into household waste; they should be disposed ofpursuant to statutory provisions.Specimens should also be disposed of according to good professional practice, observing boththe applicable statutory provisions and in-house work instructions.