CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICECarl Zeiss Trouble shooting Axio Examiner86 M60-2-0003 e 05/2012Problem Cause CorrectionThe 12 V 100 W halogen lampdoes not light even though theon/off switch is switched on.Power plug not plugged into themains outlet.Plug the power plug into the mainsoutlet, checking both equipmentvoltage and line voltage.The 12 V 100 W halogen lamp is notinstalled.Insert he 12 V 100 W halogen lamp,see page 37.The 12 V 100 W halogen lamp isdefective.Replace the 12 V 100 W halogenlamp, see page 39.The mandatory 12 V 100 W halogenlamp is not used.Use the mandatory 12 V 100 Whalogen lamp.Fuses are defective. Replace the fuses, see page 83.The 12 V 100 W halogen lamp isflickering. Its luminosity is notstable.The 12 V 100 W halogen lamp is atthe end of its mean service life.Replace the 12 V 100 W halogenlamp, see page 39.AC power cord not installed correctlyor broken.Connect the AC power cordcorrectly, or replace it.The pins of the 12 V 100 W halogenlamp are not inserted correctly intothe socket.Insert the pins of the 12 V 100 Whalogen lamp correctly into thesocket,see page 39.The motor-operated lowermicroscope part does not react.The microscope is not connected to itscontrol box.Connect the microscope,see page 46.The control box is either not switchedon or not connected.Connect or switch on the controlbox, see page 46.The docking station is not connected. Connect the control cable,see page 46.The motors are switched off. Check whether the blue indicatorlight is lit (Figure 1-4/19); if not,switch on the motors again,see page 73.