START-UPAxio Examiner Establishing a mains connection, and switching … Carl ZeissM60-2-0003 e 05/2012 47For connecting examiner control and docking station to the Axio Examiner mot., please proceed asfollows:• Connect both sockets at the lower microscope part to the sockets (Figure 3-29/4 and 5) at theexaminer control unit.• Connect the sockets at the upper microscope part to the sockets (Figure 3-29/2) at the examinercontrol unit.• If required, connect the docking station to the examiner control unit (Figure 3-29/3).All connections at the Axio Examiner mot. have a mechanically coded design. Do not use excessiveforce when connecting the units.• Connect the HAL 100 three-pole illuminator plug into the corresponding three-pole jack at theexaminer control unit.3.2.2 Connecting and switching the HAL100 halogen lamp on or offThe HAL 100 halogen lamp is powered andswitched either on or off by the external powersupply unit.The external power supply unit may be connectedto a line voltage of between 100 and 240 VAC,50 ... 60 Hz. The power supply unit will automati-cally adapt to the corresponding line voltagewithin that range.• Plug the three-pole HAL 100 illuminator pluginto the separate external power supply unit(Figure 3-30/2).• Connect the AC power socket of the externalpower supply unit (Figure 3-30/3) to a mainsoutlet using an AC power cord.• Press the power button (Figure 3-30/1) toswitch the HAL 100 halogen lamp either on oroff.When attaching the HAL 100 halogen lamp to the transmitted light socket, connect theillumination control cable of the external power supply unit to the stand (Figure 3-18/4).Figure 3-30 External power supply unit forHAL 100(front and rear side)