START-UPAxio Examiner Installing standard components Carl ZeissM60-2-0003 e 05/2012 433.1.15 HBO 100 illuminatorInserting the HBO 103 W/23.1.15.1mercury vapor short arc lampFor safety reasons, the HBO 100 illuminator andthe HBO 103 W/2 mercury vapor short arc lampwill be delivered to the customer each in aseparate package.Accordingly, the first step before starting up thisilluminator is to insert the HBO 103 W/2 lamp intothe illuminator casing.The insertion or replacement of the HBO 103 W/2lamp is described in the operating manual whichaccompanies the device.CAUTIONUse an FL attenuator for changingtransmission (unless permanentlyinstalled). Neutral-density filters will notlast in the long run.Attaching the HBO 1003.1.15.2illuminator• Remove the cap from the reflected light socket(Figure 3-25/1).• Insert the illuminator casing (Figure 3-25/3) withthe dovetail ring into the reflected light socket(Figure 3-25/1) at the unit's rear side, andsecure by a clamping screw (Figure 3-25/2)using a ball head-screwdriver (width A/F 3).• Connect the multipolar illuminator plug of theHBO 100 illuminator to the appliance bushing(Figure 3-26/1) of the HBO 100 W transformer,and secure it by a retaining ring.• Connect the AC power cord first to the ACpower inlet (Figure 3-26/2) of the HBO 100 Wtransformer, and then to a mains outlet.Figure 3-25 Attaching the HBO 100 illuminatorFigure 3-26 HBO 100 W power supply unit