Axioskop 2 plus START-UPAxioskop 2 mot plus Initial start-up Carl ZeissB 40-075 e 10/03 2-92.1.10 HAL 100 halogen illuminator(1) Attachment of HAL 100 halogenilluminator• Remove cover from transmitted-light tube(2-11/3).• Insert dovetail 2-12/2) of lamp housing (2-11/1)in transmitted-light tube (2-11/3) on theinstrument rear and use the SW 3 ball-headedscrewdriver to tighten clamping screw (2-11/2).• Connect 3-pin lamp plug (2-11/5) to 3-pin12 V/100 W socket (2-11/4) on the instrumentrear.(2) Adjustment of HAL 100 halogenilluminatora) Coarse adjustment• Loosen clamping screw (2-11/2) and remove thehalogen illuminator from the microscope stand.• Switch on microscope as described under 0.• Direct the light beam against a projection area(wall) approx. 3 m away.Do not look into the light exit openingof the illuminator.• Set adjusting screw (2-12/3) using SW 3 ball-headed screwdriver in such a way that bothimages of the lamp coil are as sharply focusedas possible on the projection screen.• Then set adjusting screws (2-12/4 and 5) in sucha way that the lamp coil of one image exactlyfills the gaps in the reflector image(2-12/1).Fig. 2-11 Attachment of HAL 100 halogenilluminatorFig. 2-12 Adjustment of HAL 100 halogenilluminator