START-UP Axioskop 2 plusCarl Zeiss Attachments and conversions Axioskop 2 mot plus2-18 B 40-075 e 10/032.2.5 Attachment of D polarizer• Hold the polarizer (2-27/6) parallel to thebottom of the condenser carrier (2-27/3) andscrew in the fastening bolts (2-274) of thepolarizer (2-27/6) with angled adjusting lever(2-27/5) to the front threaded hole on the leftbelow the condenser carrier (2-27/3) as far asthe stop.• Screw in the stop bolt (2-27/2) with theadjusting lever (2-27/1) as far as the stop in therear threaded hole of the condenser carrier(2-27/3)2.2.6 Attachment of SENARMONT DICpolarizer• Hold polarizer (2-28/6) parallel to the undersideof the condenser carrier (2-28/3) and screw stopbolt (2-28/4) of the polarizer (2-28/6) into thethreaded hole on the left below the condensercarrier (2-28/3) until stop using the angledadjusting lever (2-28/5).• Use adjusting lever (2-28/1) to screw stop bolt(2-28/2) into the rear threaded hole of thecondenser carrier (2-28/3) until stop.Fig. 2-27 Attachment of D polarizerFig. 2-28 Attachment of SENARMONT DICpolarizer