START-UPAxiovert 200 Attachment of microscope stages Carl ZeissB 40-080 e 03/01 2-11Because of the large travel range ofthe scanning stage, it may happenthat the stage frame collides withthe objectives at the end of thestage travel.2.5.3 Attachment of specimen stage250x230, object guide and mount-ing frame for object guide (M)The specimen stage is attached to the contactpoints of the stand using a spacer bar and a spacerdisk.• Use the two shorter hexagonal screws to screwthe spacer bar (2-10/5) to the two front contactpoints.• Place spacer disk (2-10/3) on the rear contactpoint.• Place specimen stage (2-10/2) on the stand andfirst fix it to the rear contact point from aboveusing the longer hexagonal screw (2-10/1).Make sure that the screw goes through thedrilled hole of the spacer disk.• Then screw specimen stage to the right and leftof the spacer bar from below using twohexagonal screws (2-10/4).• Also tighten the rear screw (2-10/1).• Attach object guide (2-11/1) to the specimenstage from the right or left and fix it in positionfrom below using three hexagonal screws(2-11/2).• Then push mounting frame for object guide (M)(2-11/3) under the two springs of the objectguide from below until it clicks into position.Fig. 2-10 Attachment of specimen stage250x230Fig. 2-11 Attachment of object guide andmounting frame