OPERATIONCarl Zeiss Axiovert 200 M Axiovert 2003-22 B 40-080 e 03/01Motorized 5-position reflector turret (3-13/3)− Accepts the maximum of 5 reflector modules for epi-fluorescence.− Fast change of reflector position through pressing the key (forward or backward) on the right of thestand (3-13/5).− The activated reflector position is marked with a line on the right of the reflector turret. The reflectorturret can also be operated manually.Key REFLECTOR (3-13/5)− Switches the reflector turret forwards or backwards by one position. The reflector turret can also beoperated manually.− Key pressed briefly: current position is shown in the LCD display for 4 s (or until the next pressof the key);without LCD display, the position is approached immediatelykey pressed again: target position is displayed and approached− If a position with a filter set is in the beam path, the filter set is shown in the lower line of the LCDdisplay. The voltage of the halogen illuminator is displayed only as a digit. If the halogen illuminator isswitched off, 0 V is displayed. If no filter set has been configured for the reflector position, thereflector position is displayed. However, the filter set and the reflector position are only displayed ifthe FL fluorescence shutter is open (FL is displayed in the upper line of the LCD display).FL key on/off (3-13/10)− Switches the fluorescence shutter on or off alternately.− pressed briefly (< 1 s): FL shutter open (FL appears in the LCD display) or closed.− pressed long (> 2 s): background illumination of the LCD display is switched on or off.− If the key is kept pressed when the microscope is switched on, the Light Manager is deactivated. TheLight Manager is activated automatically when the microscope is switched on the next time.